Model R4815/R4915 Remote Airborne Particle Counters

• 0.5 and 5.0 microns
• 1.0 cfm flow rate
• 2 size channels
• RS485 serial communications (R4815)
• 4–20 mA output (R4915)

Connecting remote particle counters to your computer (Model R4815) or SCADA (Model R4915) system has never been easier than with Met One’s miniature particle counters. These counters can be mounted at many locations around the cleanroom. All you need to provide is a vacuum source, DC power and a communication line.

The Model R4815 uses the standard RS485 serial communications, allowing instrument-to-instrument wiring up to a total length of 4000 feet (1220 m). The counter will respond to simple ASCII command instructions including commands to START count, STOP count, REPORT count, send location number, etc. The information reported includes the status of the counter along with the particle counts in two size ranges. For a graphical or spreadsheet display of data, add on the continuous monitoring features of Particle Vision® Online Software.

The Model R4915 uses the industry standard 4–20mA interface. Loop current is driven by voltages between 15 and 30 VDC. The sensor uses an RS485 serial interface for programming count time, hold time, alarm level, alarm delay, and scaling for the
4–20 mA output. The programming is retained through power off and on cycles.

For both models, a user-programmed alarm level can be set for high counts on the first channel. When high counts trigger an alarm, the LED on the counter flashes and an alarm signal is set at the connector. The alarm will automatically reset at the start of the next count cycle. The stainless steel enclosure is compatible with most process environments. There is no fan to disturb airflows, allowing the counters to be placed close to sensitive process areas.

Specifications

Size Channels (µm) 0.5, 5.0

Flow Rate 1.0 cfm (28.3 L/min)

Flow Control Critical orifice; requires 18" Hg (450 mbar) vacuum minimum

Light Source Laser diode

Coincidence Loss 5% at 400,000 particles/ft3

Inlet Pressure Ambient to 0.1" Hg vacuum

Indicators Power and Sensor/Alarm LEDs

Power 6 VDC (±10%) at <125 mA

Alarm Output Open collector FET maximum sink current (on) 150 mA;

Maximum DC voltage (off) 30 VDC

Connector DB-15 (female)

Weight 10.7 oz (0.3 kg)

Tubing Connections Inlet 1.0 cfm, 1/4" I.D., Outlet 1/4" I.D.

Environment Operating 55 to 84°F (12 to 29°C), 20 to 95% relative humidity, non-condensing

Storage -40 to 160°F (-40 to 70°C), up to 98% relative humidity, non-condensing

Accessories Included Isokinetic Probe; DB-15 Connector; Operator’s Manual

When ordering, specify: Model R4815 or Model R4915

Scaling for 4–20 mA output (Model R4915)

Optional Accessories

Switching Power Supply

Power Adapter

Wall Plate Power Adapter

Vacuum Pump

Sample or Vacuum Tubing

Particle Vision Online Software