| 6390
deadweight tester.
|
 |
- Range:
-1 bar to 40 bar, -15 to 600 psi (kg/cm² also
available)
Higher accuracy :0.01% of reading.
Standard accuracy: 0.025% of reading.
Traceability: NPL, NIST.
Weights: Stainless steel, non magnetic.
System pressure: Visual indicator.
System source: External positive and or vacuum
supply required.
System control: Fine adjustment controls. Two
test stations.
Weight: Tester 9kg - Weights:Pressure 40kg
/Vacuum 1.25kg
Dimensions: 455 x 340 x 400mm.
Standard corrections: Gravity: 9.80665 m/s²,
or local gravity on request /Temperature: 20°C
/ Air density 1.2 kg/m³
|
|
|
|